Maintenance Case For FUJI EG-250WR5 Video Gastroscope

2023-10-25

latest company case about Maintenance Case For FUJI EG-250WR5 Video Gastroscope

Equipment Model FUJI EG-250WR5 Video Gastroscope

 


Test Results

 


1, The image can not display.
2, After replacing the drive plate, there is still no image in the test. CCD has problems.
3,10% of wires inside the light beam were broken.
4, The insertion tube is ok.
5, Biopsy channel was worn.
6, The wire was worn.
7, Light guide bundle has aging conditions and cracks.

 


Maintenance Program

 


Replacement:CCD; Light Bundle; Biopsy Channe; Water Channel; Light Guide Tube; Wire; Bending Rubber

 


Repair Cycle

 


7 working days.

 


Recommendations For Maintenance And Care

 


1, Regularly use a leak detector to measure the leak (the area between inflating and red); If notice that the air leaks, send the item for maintenance in time.
2, It needs special management, regular maintenance, inspection, repair, and record.

 