2026-11-10
Many operating theatres still rely on the Olympus UHI-2 insufflator because it is simple, predictable and well understood by surgical teams. Managing a legacy fleet well is mostly about verification discipline: measuring actual delivered pressure and flow instead of trusting the display.
![]()
Stock photo: UHI-2 endoscopic insufflator with maximum flow and volume indication.
The unit offers continuous flow control, a maximum flow and volume indicator, and an integrated overpressure release path. For routine laparoscopy these functions cover the clinical need, and the controls are familiar to nursing staff who rotate between theatres.
![]()
Control panel and gas outlet detail.
A practical schedule is a functional check before each list, a full verification every six months, and a preventive service annually that includes internal inspection, tubing replacement where applicable and calibration against a reference gauge. Document every measurement; trends reveal drift long before a fault appears.
The main lifecycle risk is not the pump but the electronics and the front-panel assembly. Keep at least one serviceable spare unit per hospital site, and hold consumable items such as tubing, filters and seals in stock. When board-level parts become unavailable, plan replacement rather than reactive repair.
Replace when pressure accuracy cannot be restored by calibration, when the relief valve fails verification, or when two or more repairs occur within a year. At that point the total cost of ownership usually exceeds the cost of a newer unit with digital documentation.
Only with tubing approved for the unit and gas path. Confirm compatibility with the manufacturer's accessories list before use.
At minimum every six months, and after any repair or transport between sites.
Need verified UHI-2 units or service support? Contact our team for availability and test records.