Insufflation is the part of laparoscopy that everyone notices when it goes wrong. The Olympus UHI-3 was designed around that reality: maintain the set pressure with enough flow reserve to recover quickly from leaks, specimen retrieval and instrument exchange.
![]()
Stock photo: UHI-3 insufflator in our inventory.
Two numbers describe an insufflator's behaviour: the maximum flow it can deliver and the pressure it will not exceed. High flow without accurate pressure control is unsafe; accurate pressure without flow reserve means the field collapses every time a trocar leaks. The UHI-3 balances both, which is why it is often chosen for longer procedures.
![]()
Front panel with pressure and flow controls.
Verify displayed pressure against an external gauge at several set points, confirm the overpressure relief path operates correctly, and check that flow remains stable with a controlled leak. Replace tubing and filters at the interval specified for the unit, and keep the gas outlet free of debris.
For short routine cases, a well-maintained UHI-2 remains adequate. For long laparoscopic and robotic-assisted lists where field stability directly affects duration, the flow reserve of the UHI-3 usually justifies the investment.
Not when the pressure limit is verified and the relief path works. Flow and pressure limiting are separate control functions.
Approved insufflation tubing, a gas filter where specified, and a verified CO2 supply with adequate regulator performance.
Need a UHI-3 unit or accessories? Contact our team for stock and test information.